Interface 5030P

Key features include:

  • Maximum Applied Potential - ±6 V
  • Maximum Current - +/- 5A
  • Measure both half cells simultaneously
  • EIS - 10 μHz - 20 kHz
What it does
  • Built for single-cell testing of batteries, fuel cells, electrolyzers and other energy devices
  • Charge, discharge and cyclic charge-discharge, plus cell voltage, self-discharge and leakage rate
  • Six current ranges from 50 µA to 5 A, up to ±5 A
  • Built-in EIS to 20 kHz
  • Second electrometer for a second voltage input
  • ±6.5 V compliance at up to ±5 A

系统特点:

Interface 5000P

The Interface 5000P is designed for researchers studying different energy storage and conversion tests. It includes simultaneous measurement of voltage and impedance of two half-cells and full-cells, etc. Not only saves time, but can also characterize full cells in the same experiment.

Characterize two half-cells

The Interface 5000P can measure two half-cells simultaneously during the experiment. With an embedded reference electrode, the dual electrometer measures the potential of the anode and cathode simultaneously, even in full cells.

dual electrometers for monitoring both anode and cathode, in addition to the full
(Click image to Enlarge)

Upgradability

The Interface 5000P can be upgraded to the advanced test system Interface 5000E. A series of electrochemical tests can be performed. For inquiries on field upgrades, contact GAMRY.

更多细节

Below are the features of the Interface 5000P Potentiostat.

  • Electrochemical energy - including charge, discharge, cycle charge and discharge, potentiostatic, galvanostatic, self-discharge, leakage current, read battery voltage
  • Electrochemical Impedance Spectroscopy - including performing galvanostatic EIS up to 200,000 Hz.

Electrochemical Impedance Spectroscopy - including performing galvanostatic EIS up to 200,000 Hz.

  • 2, 3, 4 and 5 electrode measurement
  • Measure both half-cell voltage and full-cell voltage
  • Electrical isolation
    •   Floating: For autoclaves, mechanically stressed equipment or pipelines.
  • portable
    • The size of a chemistry textbook, weighing only 2.5 kg. USB 2.0 connection to Windows computer.
  • Built-in EIS
    • EIS measurements from 10 μHz to 20 kHz using advanced DDS.
  • DSP (Digital Signal Processing) Mode
    • Oversampling mode improves signal-to-noise ratio and accurate capacitance measurements. .
  • Current Interruption IR Compensation

    • The Gamry potentiostat and its control software use linear regression to accurately measure and achieve IR compensation.
  • Auxiliary I/O
    •      Control third-party devices through I/O design: realize external signal input, analog voltage output, analog current output, auxiliary A/D input and digital I/O connector.

    • RTD temperature probe measurement
  • hedging period
    •  2 year factory warranty.
    • Published EIS accuracy, not claimed accuracy
  • Interface 5030P The Gamry Interface 5030P is designed for people doing cell-level testing of various energy storage and conversion devices.  It even includes capabilities for monitoring both half cell voltages in addition to the whole cell voltage during an experiment. Imagine the time savings in being able to characterize both half cells and the full cell in the same experiment.  

    Characterizing Both Half Cells

    The Interface 5000P has the ability to monitor both half cells during experiments. Two electrometers measure the potential of the anode and cathode in addition to the full cell voltage when used with an embedded reference electrode.

    dual electrometers for monitoring both anode and cathode, in addition to the full
    Interface 5000 ACP
    Upgradeabilty

    The Interface 5030P also comes in a more advanced model, the Interface 5030E.  This model provides full capabilities for all application areas.  An in the field upgrade is available to upgrade an Interface 5000P to an Interface 5000E.  Contact us for details regarding upgrades.

下栽

参数与

system

Electrode connection 2, 3, 4 or 5 electrodes
High Current ±5 A
Current range 6 (50 μA - 5 A)
Current range (including internal gain) 8
Small current resolution 15 pA
Large applied potential ± 6V
Rise Time <1 μs
Small time base  10 μs
Noise and ripple (typical) <20 μV rms
   

Control amplifier

Trough pressure ± 8.5 / ± 2.5 V
High Current >±5 A
Speed ​​setting 5
Gain bandwidth (typical) 1050, 250, 43, 4.4, 0.5 kHz
   

EIS measurement

Frequency Range 10µHz – 20K Hz
EIS accuracy See Accuracy
Contour Plot
Applied voltage amplitude N/A
Applied current amplitude 5 The maximum
   

Potentiometer

input resistance >10 12 Ω || <2 pF
Input Current <25 pA
Bandwidth (-3 dB) (typical) >12 MHz at -3 dB
Common mode rejection ratio >98 dB (10 kHz)
>88dB (100 kHz)
   

Potential

Application accuracy    ±1 mV ±0.2% setting range
Applied resolution 200 μV, 50 μV, 12.5 μV/bit
measurement accuracy ±0.5 mV ±0.2% of reading range
Measurement resolution 200 μV, 20 μV, 2 μV/bit
   

Current

Application/measurement accuracy ±25 pA ±0.05% of range ±0.2% of value
Application/measurement resolution 0.0033% full-scale/bit
bandwidth >5MHz (5mA)