We will no longer offer support or repairs for Series G potentiostats at the end of 2019. In the interim if any of these instruments returned for repairs are unable to be repaired, we may be able to offer some trade-in value towards a new potentiostat.  Please contact our Sales Department to discuss 'Trade-In' offers available to upgrade to the latest instrument.
*****
Effective May 1, 2016 we will no longer offer support or repairs on our PCI4 potentiostats.  Please contact our Sales Department to discuss current 'Trade-In' offers available to upgrade to the latest instrument.

FAQs & Trouble Shooting

The information contained here addresses specific issues, frequently asked questions and troubleshooting.  For general hardware specifications/details, go to the relevant section of the individual product pages.  For more detailed information on specific applications, see the Applications Notes section.  For any questions not addressed here, contact one of Gamry's electrochemical experts through the Technical Support Form.

installation setup

Installation & Setup

How to install your Gamry equipment.

experimential setup

Experimental Setup & Software Basics

Setting up your experiment and running software.

faqs

Frequently Asked Questions

Common questions that customers have about setup and use of our products.

troubleshooting

Troubleshooting

Having a problem during your experiment? Start here for step-by-step troubleshooting instructions.

advanced topics2

Advanced Topics

Special topics that go beyond basic setup and use of our products.

legacy support

Legacy Device Support

Information regarding PC3, PC4, PCI4, FAS 1 & 2, Series G. - Gamry Legacy devices.

ECM8 Port Number Change

Posted by gamry on July 7, 2016

Step-by-step Instructions for Changing Your ECM8 COM port Number. Applies to Framework software version 4.35 to 5.XX

Read the full article »

Troubleshooting a Failed Calibration

Posted by Gamry on July 7, 2016

A Calibration Failure Warning Appears

Read the full article »

Reference Electrode Overview and Care

Posted by gamry on July 7, 2016

Gamry sells the Saturated Calomel and Ag/AgCl Reference Electrodes.  Proper care must be taken to ensure that your electrode continues to take accurate measurements.  The tip of the electrode is a Vycor frit (clear) or electro porous glass (opaque white).  The frit allows ionic transport into the electrode.  In order to operate properly the frit must always be kept wet with electrolyte.  Reference Electrodes can be stored for short periods of time with a rubber cap installed over the frit.  A securely fitted cap can keep a frit in good working condition for about 2 months.  For a better storage method, immerse the tip of the electrode into a beaker of saturated KCl solution.  Electrodes stored in this manner will have a working frit for a very long time.

Read the full article »

MultEchem Dongle #00000 Installed

Posted by gamry on July 7, 2016

USB ports where the dongle does not engage tightly can result in Microsoft Windows detecting the dongle but not properly reading the dongle serial number

Read the full article »

How do I copy and paste graphs and data from Echem Analyst?

Posted by gamry on July 7, 2016

The Copy to clipboard button is located on the toolbar of the Chart tab when your data file is displayed in Echem Analyst. This button gives you the option to copy a Bitmap, Metafile, or raw data directly to the Windows Clipboard.

Read the full article »

What are the file types used by Gamry Software?

Posted by Gamry on July 7, 2016

This is the preferred extension for raw data acquired by Gamry Framework software. The file contains only tab-separated raw text and can be opened in any text editor or spreadsheet program.

Read the full article »

Manually changing the Port number for your ECM8 or TDC4

Posted by gamry on July 7, 2016

When installing a second ECM8, or switching to a USB-to-Serial adapter for your ECM8 or TDC4, it is necessary to manually change the Port number in the Gamry.INI file. The correct Port number allows the Framework Software to communicate properly with the Gamry Hardware device. The INI file is accessible from the Windows Start Menu under Gamry / Framework / Gamry.ini. The file will open in your default text editor program.

Read the full article »

Installation of a MultEchem System

Posted by gamry on July 7, 2016

MultEchem systems allow multiple instruments to share a single software license on a single computer.  The system utilizes a USB dongle to activate the software license across all instruments.  For all Framework versions before 5.63 one must select “MultEchem” from the Select Features page of the installation program to install the driver for the USB dongle.  This driver is installed automatically in current versions.

Read the full article »

IC8 Becomes Unresponsive Intermittently

Posted by gamry on July 7, 2016

General instability in the operation of the IC8 can be caused by many things. Instability refers to a loss of keyboard and mouse input, loss of video or a Microsoft BSOD that can only be resolved by restarting the computer.

Read the full article »

Field Powering Your USB Potentiostat

Posted by gamry on July 7, 2016

A number of users bring potentiostats into the field and need a reliable source of short term power for experiments.  All of our USB potentiostats run from 120V AC wall outlets, but with varying power requirements.

Read the full article »