Detection and sizing of single droplets flowing in a lab-on-a-chip device by measuring impedance fluctuations

By Nour Eddin Yakdi and Fran
Published in Sensors and Actuators B: Chemical NULL 2016

Abstract

This paper is devoted to the evaluation of the electrochemical noise technique consisting in measuring the fluctuations of the electrolyte resistance (ER) between two metallic electrodes immerged in a conductive electrolyte to detect and characterize single particles circulating in a microfluidic device, without the help of optical measurements that require good visibility of the detection region. Numerical simulations were performed with the finite element method to study the influence of the dimensions of the channel and the electrodes on the ER. Measurements of the \ER\ variations due to the passage of oil droplets and plugs passing between the electrodes were carried out. Excellent agreement was obtained between the theoretical and experimental \ER\ transients, which allowed the velocity and diameter of the oil droplets to be estimated with an accuracy of a few percents in the case of droplet diameters ranging from 60 to 100 ?m. According to the numerical simulations and the amplitude of the background noise, oil droplets of diameter larger than 20

Read Article » Back